| General Specifications |
|
Electrical safety class
|
I
|
|
IP class
|
21
|
|
Applied part type
|
B (gamma probe)
|
|
Dimensions (h x b x d)
|
1981 mm x 598 mm x 1620 mm
|
|
Power cable length
|
5 m
|
|
Power cable plug
|
NEMA 5 (Type B) / CEE 7/7 (Type E/F)
|
|
Power supply
|
115/230 VAC ± 10 %, 50/60 Hz
|
|
Power consumption
|
1000 Wp-p (when switching on) 450 VArms (steady operation)
|
|
Max. output impedance of mains
|
5 Ω
|
|
Fuses of power supply
|
2x 3.15A, T class fuses (UL, CSA or VDE/EN approved only)
|
|
Operating temperature
|
10 – 30 °C (30 – 86 °F)
|
|
Operating relative humidity
|
30 – 70 %
|
|
Operating atmospheric pressure
|
75 – 106 kPa
|
|
Weight with all components
|
ca. 69 kg
|
|
Handle (sterile) length
|
100 mm
|
|
Handle (sterile) diameter
|
32.7 mm
|
|
Handle (unsterile) length
|
96 mm
|
|
Handle (unsterile) thickness
|
34 mm
|
|
Lens aperture of optical camera
|
F1.4 – infinity (adjustable)
|
|
Focal length of optical camera
|
4 mm / 8 mm
|
|
Optical camera resolution
|
1032 x 778
|
|
Camera sensor size
|
1/3"
|
|
Optical video frame rate
|
30 fps
|
|
Color depth of optical image
|
12 bit
|
| Image Specifications |
|
Sampling frequency
|
20 Hz
|
|
Reconstruction time
|
30 – 90s
|
|
Voxel size
|
5 mm x 5 mm x 5 mm
|
|
Min. spatial resolution
|
5 mm x 5 mm x 5 mm
|
|
Max. volume of reconstruction
|
50 cm x 50 cm x 50 cm
|
|
Data transfer interface
|
Data transfer over encoded USB stick / Ethernet / WLAN
|
|
Data storage
|
DICOM images, PDF report, JPEG screenshots and optional AVI video
|
|
declipseSPECT is compatible with several gamma probes and the specifications of the nuclear detector will
vary accordingly to the gamma probe being used. For more details on the nuclear detector, system positioning and accessories specifications or if you want to know if your gamma probe is compatible with the declipseSPECT email here to This e-mail address is being protected from spambots. You need JavaScript enabled to view it
|